VOC Test Chamber System 15

VOC Emission Test Chambers 13_14_15


Strict Standards: Only variables should be assigned by reference in /home/clritali/public_html/templates/yoo_drive/warp/systems/joomla/layouts/com_content/article/default.php on line 143

Forbo 636 534

System 15: System to store samples for VOC analysis avoiding cross contamination; 12 independent chambers, with 87,5 L volume each; control of relative humidity of the carrier gas at the inlet of the system, operated by the Customer; control of temperature in the chambers operated by the Customer; one air outlet on the port of each chamber; 12 flowmeters, local inducation, manual flow adjustment.

This system has been designed for a French Company manufacturing flooring and acoustic insulation systems made of PVC.

Main technical features

12 independent test chambers, with 87,5 L volume each

control of relative humidity of the carrier gas at the inlet of the system, operated by the Customer in order to obtain the desired RH value inside the storing chambers

control of temperature in the chambers operated by the Customer through the control of the temperature in the room where the system is installed

separate air flowrate control on each chamber, massic flowmeters, nominal flowrate range:
12 ÷600 NL/h, accuracy ±2% F.S.

System overall dimensions: 2438 (H) x 1600 (W) x 670 (D) mm

Test chambers: electrochemically polished stainless steel, anodized aluminium blind ports,
inner dimensions 320 (H) x 560 (W) x 547 (D) mm inner volume approx. 87,5 L

Connections for air outlet: 1 stub on each port, 6 mm external diameter, 5 mm internal diameter, suitable as flexible hose connection, with adjustable insertion depth